157/159 UV-VIS Detectors




Product Resources

A list of available product resources are listed below. Click a resource category to expand the list.



Product Details

Specifications

Click Here for the 157 UV-VIS Detector Specifications

Click Here for the 159 UV-VIS Detector Specifications

 

157 UV-VIS Detector

Specifications

 

Detector Type        

Variable, Single Wavelength

Wavelength Range

190-750 nm

Lamps

Deuterium (D2) lamp with integrated GLP chip, warranted for 2000 hours

Spectral Bandwidth

11 nm at Hα line (FWHM)

Drift

3.0 x 10-4 AU/h at 254 nm (ASTM E1657-98)

Linearity

> 2.0 AU at 270 nm (ASTM E1657-98)

Wavelength Accuracy

± 2.5 nm

Wavelength Precision

0.3 nm (ASTM E1657-98)

Noise

±1.5 x 10-5 AU at 254 nm (ASTM E1657-98)

Time Constants

0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s

Flow Cell Assemblies

The following quartz flow cells are available:

Flow Cell

Path

Volume

Pressure

Analytical

10 mm

10 µL

300 bar

Analytical (PEEK)

3 mm

2 µL

30 bar

Preparative

0.5 mm

3 µL

200 bar

Preparative

                               0.5-2 mm

-

200 bar

Front Panel

Touchscreen TFT 2.4"

Communication Interface

Inputs:     Error, Start, Autozero, 1 - 10 V Analog

Outputs:  Events 1 - 3, + 5 V, 24 V Valve

(1) Analog outputs 0-5 V Scalable

Software Control

Computer control via Ethernet communication and TRILUTION® LC Software

Data Rate

50 Hz (LAN, Analog)

Environmental Conditions

Indoor use
Altitude: up to 2000 m
Temperature range: 4 - 40 °C, 39.2 - 104 °F

Humidity: below 90 % noncondensing

Power Requirements       

Frequency: 50–60 Hz
Voltage: 110–240V

Dimensions (w x d x h)

24.2 x 39.9 x 16.9 cm (9.7 x 16.0 x 6.8 in.)

Weight

5.3 kg (11.7 lbs.)

Manufacturing Standards (English)

Meets applicable safety and EMC certification standards; UL and CE certified

 

 

159 UV-VIS Detector

Specifications

 

Detector Type        

Variable, Multiple Wavelength

Data Channels

4 Wavelengths simultaneously

Wavelength Range

190-750 nm

Lamps

Deuterium (D2) lamp with integrated GLP chip, warranted for 2000 hours

Spectral Bandwidth

6 nm at Hα line (FWHM)

Drift

1.5 x 10-4 AU/h at 254 nm (ASTM E1657-98)

Linearity

> 2.2 AU at 270 nm (ASTM E1657-98)

Wavelength Accuracy

± 2.0 nm

Wavelength Precision

0.4 nm (ASTM E1657-98)

Noise

±0.75 x 10-5 AU at 254 nm (ASTM E1657-98)

Time Constants

0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s

Flow Cell Assemblies

The following quartz flow cells are available:

Flow Cell

Path

Volume

Pressure

Analytical

10 mm

10 µL

300 bar

Analytical (PEEK)

3 mm

2 µL

30 bar

Preparative

0.5 mm

3 µL

200 bar

Preparative

                             0.5-2 mm

-

200 bar

Front Panel

Touchscreen TFT 2.4"

Communication Interface

Inputs:     Error, Start, Autozero, 1 - 10 V Analog

Outputs:  Events 1 - 3, + 5 V, 24 V Valve

               (2) Analog outputs 0-5 V Scalable

Software Control

Computer control via Ethernet communication and TRILUTION® LC Software

Data Rate

80 Hz (LAN, Analog)

Environmental Conditions

Indoor use
Altitude: up to 2000 m
Temperature range: 4 - 40 °C, 39.2 - 104 °F

Humidity: below 90 % noncondensing

Power Requirements       

Frequency: 50–60 Hz
Voltage: 110–240V

Dimensions (w x d x h)

24.2 x 39.9 x 16.9 cm (9.7 x 16.0 x 6.8 in.)

Weight

5.3 kg (11.7 lbs.)

Manufacturing Standards (English)

Meets applicable safety and EMC certification standards; UL and CE certified