PIPETMAN Neo® Mehrkanal




Specifications

 

The Pipetting Standard With Improved Comfort!

Gilson's standards are even more stringent than those required by governing bodies: On average, PIPETMAN permissible errors are 2 times better than those recommended by ISO 8655!

  • Our pipette manufacturing plant has its own accredited calibration laboratory to control the accuracy and precision of each and every Gilson pipette.
  • Our pipettes are individually calibrated and delivered with a certificate of conformity. Each pipette bears an engraved identification number for traceability.

Volumetric specifications

 Model Ordering reference Volume (µL) Gilson Maximum
Permissible Errors
ISO-8655
Permissible Errors
Systematic error (µL) Random error (µL) Systematic error (µL) Random error (µL)
8x20 F14401 2 ±0.10 ≤0.08 ±0.40 ≤0.20
10 ±0.20 ≤0.1 ±0.40 ≤0.20
20 ±0.40 ≤0.2 ±0.40 ≤0.20
12x20 F14402 2 ±0.10 ≤0.08 ±0.40 ≤0.20
10 ±0.20 ≤0.1 ±0.40 ≤0.20
20 ±0.40 ≤0.2 ±0.40 ≤0.20
8x200 F14403 20 ±0.50 ≤0.25 ±3.20 ≤1.20
100 ±1.00 ≤0.40 ±3.20 ≤1.20
200 ±2.00 ≤0.60 ±3.20 ≤1.20
12x200 F14404 20

±0.50

≤0.25 ±3.20 ≤1.20
100 ±1.0 ≤0.40 ±3.20 ≤1.20
200 ±2.00 ≤0.60 ±3.20 ≤1.20

 Technical Specifications

  • Stainless steel piston.
  • PVDF handle.

 Click here to see PIPETMAN Neo Single Channel Pipette Options