PIPETMAN® G




Specifications

 

Premium Pipetting Comfort!

Gilson's standards are even more stringent than those required by governing bodies: On average, PIPETMAN permissible errors are 2 times better than those recommended by ISO 8655!

  • Our pipette manufacturing plant has its own accredited calibration laboratory to control the accuracy and precision of each and every Gilson pipette.
  • Our pipettes are individually calibrated and delivered with a certificate of conformity. Each pipette bears an engraved identification number for traceability.

Volumetric specifications

 Model Ordering reference Volume tested
 (µL)
Gilson Maximum
Permissible Errors
ISO-8655
Permissible Errors
Systematic error (µL) Random error (µL) Systematic error (µL) Random error (µL)

P2G

F144054P
F144054M
0.2 ±0.024 ≤0.012 ±0.08 ≤0.04
0.5 ±0.025 ≤0.012 ±0.08 ≤0.04
2 ±0.030 ≤0.014 ±0.08 ≤0.04

P10G

F144055P
F144055M
1 ±0.025 ≤0.012 ±0.12 ≤0.08
5 ±0.075 ≤0.030 ±0.12 ≤0.08
10 ±0.100 ≤0.040 ±0.12 ≤0.08

P20G

F144056P F144056M 2 ±0.10 ≤0.03 ±0.20 ≤0.10
10 ±0.10 ≤0.05 ±0.20 ≤0.10
20 ±0.20 ≤0.06 ±0.20 ≤0.10

P100G

F144057P F144057M 10 ±0.35 ≤0.10 ±0.80 ≤0.30
50 ±0.40 ≤0.12 ±0.80 ≤0.30
100 ±0.80 ≤0.15 ±0.80 ≤0.30

P200G

F144058P F144058M 20 ±0.50 ≤0.20 ±1.60 ≤0.60
100 ±0.80 ≤0.25 ±1.60 ≤0.60
200 ±1.60 ≤0.30 ±1.60 ≤0.60

P1000G

F144059P F144059M 100 ±3 ≤0.6 ±8 ≤3.0
500 ±4 ≤1.0 ±8 ≤3.0
1000 ±8 ≤1.5 ±8 ≤3.0

P5000G

F144066 500 ±12 ≤3 ±40 ≤15
2500 ±15 ≤5 ±40 ≤15
5000 ±30 ≤8 ±40 ≤15

P10mL G

F144067 1000 ±30 ≤6 ±60 ≤30
5000 ±40 ≤10 ±60 ≤30
10000 ±60 ≤16 ±60 ≤30

 

Test tube