Volumetric specifications

Single Channel Models

  Model Ordering reference Volume (µL) Gilson Maximum Permissible Errors ISO-8655 Permissible Errors
      Systematic error (µL) Random error (µL) Systematic error (µL) Random error (µL)
F2 F123770 2 ±0.08 ±0.03 ±0.08 ≤0.04
F5 F123771 5 ±0.10 ≤0.04 ±0.125 ≤0.075
F10 F123772 10 ±0.10 ≤0.05 ±0.12 ≤0.08
F20 F123604 20 ±0.20 ≤0.06 ±0.20 ≤0.10
F25 F123775 25 ±0.25 ≤0.07 ±0.50 ≤0.20
F50 F123778 50 ±0.40 ≤0.15 ±0.50 ≤0.20
F100 F123784 100 ±0.80 ≤0.25 ±0.80 ≤0.30
F200 F123605 200 ±1.60 ≤0.30 ±1.60 ≤0.60
F250 F123787 250 ±3.00 ≤0.75 ±4.00 ≤1.50
F300 F123788 300 ±3.50 ≤0.75 ±4.00 ≤1.50
F400 F123789 400 ±3.60 ≤0.80 ±4.00 ≤1.50
F500 F123790 500 ±4.00 ≤1.00 ±4.00 ≤1.50
F1000 F123606 1000 ±8.00 ≤1.30 ±8.00 ≤3.00
F5000 F123607 5000 ±30.00 ≤8.00 ±40.00 ≤15.00

 


Technical specifications

  • Stainless steel piston.
  • PVDF handle.

 


Models & Volume Range

14 models of PIPETMAN F to cover the range from 0.2 µL to 5000 µL.

  Model Ordering reference Volume (µL)
  F2 F123770 2
  F5 F123771 5
  F10 F123772 10
  F20 F123604 20
  F25 F123775 25
  F50 F123778 50
  F100 F123784 100
  F200 F123605 200
  F250 F123787 250
  F300 F123788 300
  F400 F123789 400
  F500 F123790 500
  F1000 F123606 1000
  F5000 F123607 5000