PLC Purification Systems




Specifications

 

PLC 2050 / 2250 / 2500 Purification Systems Specifications

Flow Rate / Max Pressure

PLC 2050

  • 50 mL/min,  with pressure up to 300 bar (4351 psi)

PLC 2250

  • 250 mL/min, with pressure up to 230 bar (3336 psi)

PLC 2500

  • 500 mL/min, with pressure up to 110 bar (1595 psi)

Flow Rate Performance

  • 2% accuracy (with H2O degassed at 20°C)
  • Repeatability better than 0.5%
  • Linear response to gradients from 2% to 98%

Binary Gradient (Standard) or Quaternary (Optional) Injection

  • Automatic or manual look injection valve (optional)
  • Additional injection pump (optional)
  • Additional GX-241 Liquid Handler (optional)

Gradient Former

  • Binary (standard)
  • Quaternary (optional)

System Protection

  • Software with minimum and maximum pressures adjustable

Control

  • Touchscreen PC with Gilson Gilder Prep software

Storage

  • Up to 32 methods in software memory and 80 GB of dard disk memory

AC Mains Supply Voltage

  • 110V pt 240V. 50/60 Hz, 350 W (maximum)

Electrical Protection

  • Delayed action fuses T 6.3A H 250 V~

Power Distribution

  • TT or TN power system only, transient over-voltages accourding to installation category II, Class 1 equipment, requires mains socket plug with protective earth terminal

Valve Options

  • Automatic backflush valve (optional)
  • Manual column switching valve (optional)

Column Holder

Built-in tightening clamp for 10– 100 mm ID columns (standard)

Flash cartridges: 2.6 g – 1.2 kg (standard)

Additional standalone holder for flash/prep columns (optional):

  • Maximum height: 400 mm
  • Maximum diameter: 115 mm

Detector Options

  • UV monochromator single wavelength detector, 254 nm by default (standard)
  • UV monochromator dual wavelength detector, 254/280 nm by default (optional)
  • UV detector, 4-wavelength DAD, 200– 400 nm (optional)
  • UV/VIS detector, 4-wavelength DAD, 200– 600 nm
    (or 190– 840 nm) (optional)
  • VERITY® 1900 MS Detector (optional)
  • VERITY® 1810 Conductivity and pH Monitor (optional)
  • Built-in ELSD (optional)

Collection

  • Several rack options available
  • Rack overlays for easy tube indentification (optional)
  • Funnel rack for a large and variable collection volume in external high-capacity containers
  • Fraction collector cover kit (optional)