PLATEMASTER®




Specifications

 

Unmatched Specifications

PLATEMASTER is a high quality instrument that offers excellent accuracy and precision. The figures given in the “Gilson Maximum Permissible Errors” table (below) were obtained using PIPETMAN Tips. These figures are guaranteed only when genuine PIPETMAN Tips are used.

PLATEMASTER
Model

Volume
Tested

Maximum Permissible Error

 

Gilson

ISO 8655

 

Systematic Error (µL)

Random Error (µL)

Systematic Error (µL)

“Random Error (µL)

 

P20

Min.:

1

± 0.12

≤ 0.10

± 0.40

≤ 0.20

 

 

10

± 0.12

≤ 0.10

± 0.40

≤ 0.20

 

Max.:
 

20

± 0.20

≤ 0.18

± 0.40

≤ 0.20

 

 

P220

Min.:

2

± 0.12

≤ 0.15

± 8.00

≤ 3.00

 

 

5

± 0.25

≤ 0.175

± 8.00

≤ 3.00

 

 

20

± 0.40

≤ 0.30

± 8.00

≤ 3.00

 

 

100  

± 1.00

≤ 0.60

± 8.00

≤ 3.00

 

 

200

± 1.60

≤ 0.80

± 8.00

≤ 3.00

 

Max.:

220  

± 1.80

≤ 0.80

± 8.00

≤ 3.00

 


Complete Range Available

Model Volume Range (µL) Ordering Information
 PLATEMASTER P20 0.5 - 20 µL F110761
 PLATEMASTER P220  2 - 220 µL F110762

Recommended working temperature: 20 to 25°C

Dimensions (W x L x H) and weight: 300 x 450 x 400 mm / 16 kg

Test tube