PIPETMAN® Classic




Specifications

Gilson's standards are even more stringent than those required by governing bodies: On average, PIPETMAN permissible errors are 2 times better than those recommended by ISO 8655!

Our pipette manufacturing plant has its own accredited calibration laboratory to control the accuracy and precision of each and every Gilson pipette. Gilson pipettes are individually calibrated and delivered with a performance check report. Each pipette bears an engraved identification number for traceability.


Volumetric Specifications

Single Channel Models

 

 

 

Model

 

 

Ordering reference

 

 

Volume (µL)

Gilson Maximum
Permissible Errors
ISO-8655
Permissible Errors
Systematic error (µL) Random error (µL) Systematic error (µL) Random error (µL)
P2 F144801 0.2 ±0.024 ≤0.012 ±0.08 ≤0.04
0.5 ±0.025 ≤0.012 ±0.08 ≤0.04
2 ±0.030 ≤0.014 ±0.08 ≤0.04
P10 F144802 1 ±0.025 ≤0.012 ±0.12 ≤0.08
5 ±0.075 ≤0.030 ±0.12 ≤0.08
10 ±0.100 ≤0.040 ±0.12 ≤0.08
P20 F123600 2 ±0.10 ≤0.03 ±0.20 ≤0.10
5  ±0.10  ≤0.04  ±0.20 ≤0.10
10 ±0.10 ≤0.05 ±0.20 ≤0.10
20 ±0.20 ≤0.06 ±0.20 ≤0.10
P100 F123615 20 ±0.35 ≤0.10 ±0.80 ≤0.30
50 ±0.40 ≤0.12 ±0.80 ≤0.30
100 ±0.80 ≤0.15 ±0.80 ≤0.30
P200 F123601 50 ±0.50 ≤0.20 ±1.60 ≤0.60
100 ±0.80 ≤0.25 ±1.60 ≤0.60
200 ±1.60 ≤0.30 ±1.60 ≤0.60
P1000 F123602 200 ±3 ≤0.6 ±8 ≤3.0
500 ±4 ≤1.0 ±8 ≤3.0
1000 ±8 ≤1.5 ±8 ≤3.0
P5000 F123603 1000 ±12 ≤3 ±40 ≤15
2000 ±12 ≤5 ±40 ≤15
5000 ±30 ≤8 ±40 ≤15
P10 mL F161201 1 mL ±30 ≤6 ±60 ≤30
2 mL ±30 ≤6 ±60 ≤30
5 mL ±40 ≤10 ±60 ≤30
10 mL ±60 ≤16 ±60 ≤30

 

3 Kits Adapted to Your Application

To help you get started, two or three PIPETMAN Classic as well as useful accessories and samples of PIPETMAN DIAMOND® Tips are supplied in a single box.

Product Designation Ordering reference PIPETMAN Models DIANOND Tips
Starter Kit F167300 P20 - P200 - P1000 D200 - D1000
Micro-volume Kit F167500 P2 - P10 - P100 DFL10ST - DF100ST
Large-volume Kit F167400 P5000 - P10 mL D5000 - D10 mL

 

Technical Specifications

  • Stainless steel piston.
  • PVDF handle.