PIPETMAN® L




Specifications

 

Trusted Results

PIPETMAN® L is a high quality pipette that offers excellent accuracy and precision. The figures given in the "Gilson Maximum Permissible Errors" table below were obtained using PIPETMAN Tips. These figures are valid only when genuine PIPETMAN Tips are used.

Volumetric Specifications


Model
Volume
Tested
Maximum Permissible Errors
Gilson ISO 8655-5

Systematic
Error (µL)

Random
Error (µL)

Systematic
Error (µL)

Random
Error (µL)

P2L
(FA10001AP
or FA10001AM)

Min.: 0.2 ± 0.024 ≤ 0.012 ± 0.08 ≤ 0.04
  0.5 ± 0.025 ≤ 0.012 ± 0.08 ≤ 0.04
  1 ± 0.027 ≤ 0.013 ± 0.08 ≤ 0.04
Max.: 2 ± 0.030 ≤ 0.014 ± 0.08 ≤ 0.04

P10L
(FA10002AP
or FA10002AM)

Min.: 1 ± 0.025 ≤ 0.012 ± 0.12 ≤ 0.08
  5 ± 0.075 ≤ 0.030 ± 0.12 ≤ 0.08
Max.: 10 ± 0.100 ≤ 0.040 ± 0.12 ≤ 0.08

P20L
(FA10003AP

or FA10003AM)

Min.: 2 ± 0.10 ≤ 0.03 ± 0.20 ≤ 0.10
  10 ± 0.10 ≤ 0.05 ± 0.20 ≤ 0.10
Max.: 20 ± 0.20 ≤ 0.06 ± 0.20 ≤ 0.10

P100L
(FA10004AP
or FA10004AM)

Min.: 10 ± 0.35 ≤ 0.10 ± 0.80 ≤ 0.30
  50 ± 0.40 ≤ 0.12 ± 0.80 ≤ 0.30
Max.: 100 ± 0.80 ≤ 0.15 ± 0.80 ≤ 0.30

P200L
(FA10005AP
or FA10005AM)

Min.: 20 ± 0.50 ≤ 0.20 ± 1.60 ≤ 0.60
  100 ± 0.80 ≤ 0.25 ± 1.60 ≤ 0.60
Max.: 200 ± 1.60 ≤ 0.30 ± 1.60 ≤ 0.60

P1000L
(FA10006AP
or FA10006AM)

Min.: 100 ± 3 ≤ 0.6 ± 8 ≤ 3.0
  500 ± 4 ≤ 1.0 ± 8 ≤ 3.0
Max.: 1000 ± 8 ≤ 1.5 ± 8 ≤ 3.0

P5000L
(FA10007A)

Min.: 500 ± 12 ≤ 3 ± 40 ≤ 15
  2500 ± 15 ≤ 5 ± 40 ≤ 15
Max.: 5000 ± 30 ≤ 8 ± 40 ≤ 15

P10mL L
(FA10008A)

Min.: 1000 ± 30 ≤ 6 ± 60 ≤ 30
  5000 ± 40 ≤ 10 ± 60 ≤ 30
Max.: 10000 ± 60 ≤ 16 ± 60 ≤ 30

 

Complete Range Available

Pipette
Model
Volume
Range (µL)
Color
Code
Part Numbers
With stainless steel
ejector
With plastic
Ejector
 P2L  0.2 - 2  Orange  FA10001AM  FA10001AP
 *P10L  1 - 10  Red  FA10002AM  FA10002AP
 P20L  2 - 20  Pale Yellow  FA10003AM  FA10003AP
 P100L  10 - 100  Salmon  FA10004AM  FA10004AP
 P200L  20 - 200  Yellow  FA10005AM  FA10005AP
 P1000L  100 - 1000  Blue  FA10006AM  FA10006AP
 P5000L  500 - 5000  Purple   **FA10007A
 P10mL L  1 - 10 mL  Pale Blue   **FA10008A

*P10L model can be used up to 0.5 µL
** No tip ejector required or needed

Test tube