PIPETMAN® L
Specifications
Trusted Results
PIPETMAN, Unmatched Specifications
PIPETMAN® L is a high quality pipette that offers excellent accuracy and precision. The figures given in the “Gilson Maximum Permissible Errors” table (below) were obtained using PIPETMAN Tips. These figures are guaranteed only when genuine PIPETMAN Tips are used.
Pipette
Model |
Volume
Tested |
Maximum Permissible Errors |
| Gilson |
ISO 8655 |
| Systematic Error (µL) |
Random Error (µL) |
Systematic Error (µL) |
“Random Error (µL) |
P2L
|
Min.: |
0.2 |
± 0.024 |
≤ 0.012 |
± 0.08 |
≤ 0.04 |
| |
0.5 |
± 0.025 |
≤ 0.012 |
± 0.08 |
≤ 0.04 |
| |
1 |
± 0.027 |
≤ 0.013 |
± 0.08 |
≤ 0.04 |
| Max.: |
2 |
± 0.030 |
≤ 0.014 |
± 0.08 |
≤ 0.04 |
P10L
|
Min.: |
1 |
± 0.025 |
≤ 0.012 |
± 0.12 |
≤ 0.08 |
| |
5 |
± 0.075 |
≤ 0.030 |
± 0.12 |
≤ 0.08 |
| Max.: |
10 |
± 0.100 |
≤ 0.040 |
± 0.12 |
≤ 0.08 |
P20L
|
Min.: |
2 |
± 0.10 |
≤ 0.03 |
± 0.20 |
≤ 0.10 |
| |
10 |
± 0.10 |
≤ 0.05 |
± 0.20 |
≤ 0.10 |
| Max.: |
20 |
± 0.20 |
≤ 0.06 |
± 0.20 |
≤ 0.10 |
P100L
|
Min.: |
10 |
± 0.35 |
≤ 0.10 |
± 0.80 |
≤ 0.30 |
| |
50 |
± 0.40 |
≤ 0.12 |
± 0.80 |
≤ 0.30 |
| Max.: |
100 |
± 0.80 |
≤ 0.15 |
± 0.80 |
≤ 0.30 |
P200L
|
Min.: |
20 |
± 0.50 |
≤ 0.20 |
± 1.60 |
≤ 0.60 |
| |
100 |
± 0.80 |
≤ 0.25 |
± 1.60 |
≤ 0.60 |
| Max.: |
200 |
± 1.60 |
≤ 0.30 |
± 1.60 |
≤ 0.60 |
P1000L
|
Min.: |
100 |
± 3 |
≤ 0.6 |
± 8 |
≤ 3.0 |
| |
500 |
± 4 |
≤ 1.0 |
± 8 |
≤ 3.0 |
| Max.: |
1000 |
± 8 |
≤ 1.5 |
± 8 |
≤ 3.0 |
P5000L
|
Min.: |
500 |
± 12 |
≤ 3 |
± 40 |
≤ 15 |
| |
2500 |
± 15 |
≤ 5 |
± 40 |
≤ 15 |
| Max.: |
5000 |
± 30 |
≤ 8 |
± 40 |
≤ 15 |
P10mL L
|
Min.: |
1000 |
± 30 |
≤ 6 |
± 60 |
≤ 30 |
| |
5000 |
± 40 |
≤ 10 |
± 60 |
≤ 30 |
| Max.: |
10000 |
± 60 |
≤ 16 |
± 60 |
≤ 30 |