PIPETMAN® L Multichannel




Specifications

 

Trusted Results

PIPETMAN® L is a high quality pipette that offers excellent accuracy and precision. The figures given in the "Gilson Maximum Permissible Errors" table below were obtained using PIPETMAN Tips. These figures are valid only when genuine PIPETMAN Tips are used.

Volumetric Specifications


Model
Volume
Tested
Maximum Permissible Errors
Gilson ISO 8655-5

Systematic
Error (µL)

Random
Error (µL)

Systematic
Error (µL)

Random
Error (µL)

P8x10L
(FA10013A)

and
P12x10L
(FA100014A)

Min.: 0.5 ± 0.024 ≤ 0.012 ± 0.08 ≤ 0.04
  1 ± 0.025 ≤ 0.012 ± 0.08 ≤ 0.04
  5 ± 0.027 ≤ 0.013 ± 0.08 ≤ 0.04
Max.: 10 ± 0.030 ≤ 0.014 ± 0.08 ≤ 0.04


P8x20L
(FA1009A)
and
P12x20L
(FA10010A)

Min.: 2 ± 0.025 ≤ 0.012 ± 0.12 ≤ 0.08
  10 ± 0.075 ≤ 0.030 ± 0.12 ≤ 0.08
Max.: 20 ± 0.100 ≤ 0.040 ± 0.12 ≤ 0.08


P8x200L
(FA10011A)
and
P12x200L
(FA10012A)

Min.: 20 ± 0.10 ≤ 0.03 ± 0.20 ≤ 0.10
  100 ± 0.10 ≤ 0.05 ± 0.20 ≤ 0.10
Max.: 200 ± 0.20 ≤ 0.06 ± 0.20 ≤ 0.10


P8x300L
(FA10015A)
and
P12x300L
(FA10016A)

Min.: 20 ± 1 ≤ 0.35 ± 8 ≤ 3
 

30

150

± 1

± 1.5

≤ 0.35

≤ 0.6

± 8

± 8

≤ 3

≤ 3

Max.: 300 ± 3 ≤ 1 ± 8 ≤ 3

 

Test tube