PIPETMAN® L




Specifications

 

Unmatched Specifications

PIPETMAN® L is a high quality pipette that offers excellent accuracy and precision. The figures given in the “Gilson Maximum Permissible Errors” table (below) were obtained using PIPETMAN Tips. These figures are guaranteed only when genuine PIPETMAN Tips are used.

Pipette
Model
Volume
Tested
Maximum Permissible Errors
Gilson ISO 8655
Systematic Error (µL) Random Error (µL) Systematic Error (µL) “Random Error (µL)

P2L

Min.: 0.2 ± 0.024 ≤ 0.012 ± 0.08 ≤ 0.04
  0.5 ± 0.025 ≤ 0.012 ± 0.08 ≤ 0.04
  1 ± 0.027 ≤ 0.013 ± 0.08 ≤ 0.04
Max.: 2 ± 0.030 ≤ 0.014 ± 0.08 ≤ 0.04

P10L

Min.: 1 ± 0.025 ≤ 0.012 ± 0.12 ≤ 0.08
  5 ± 0.075 ≤ 0.030 ± 0.12 ≤ 0.08
Max.: 10 ± 0.100 ≤ 0.040 ± 0.12 ≤ 0.08

P20L

Min.: 2 ± 0.10 ≤ 0.03 ± 0.20 ≤ 0.10
  10 ± 0.10 ≤ 0.05 ± 0.20 ≤ 0.10
Max.: 20 ± 0.20 ≤ 0.06 ± 0.20 ≤ 0.10

P100L

Min.: 10 ± 0.35 ≤ 0.10 ± 0.80 ≤ 0.30
  50 ± 0.40 ≤ 0.12 ± 0.80 ≤ 0.30
Max.: 100 ± 0.80 ≤ 0.15 ± 0.80 ≤ 0.30

P200L

Min.: 20 ± 0.50 ≤ 0.20 ± 1.60 ≤ 0.60
  100 ± 0.80 ≤ 0.25 ± 1.60 ≤ 0.60
Max.: 200 ± 1.60 ≤ 0.30 ± 1.60 ≤ 0.60

P1000L

Min.: 100 ± 3 ≤ 0.6 ± 8 ≤ 3.0
  500 ± 4 ≤ 1.0 ± 8 ≤ 3.0
Max.: 1000 ± 8 ≤ 1.5 ± 8 ≤ 3.0

P5000L

Min.: 500 ± 12 ≤ 3 ± 40 ≤ 15
  2500 ± 15 ≤ 5 ± 40 ≤ 15
Max.: 5000 ± 30 ≤ 8 ± 40 ≤ 15

P10mL L

Min.: 1000 ± 30 ≤ 6 ± 60 ≤ 30
  5000 ± 40 ≤ 10 ± 60 ≤ 30
Max.: 10000 ± 60 ≤ 16 ± 60 ≤ 30

 

Complete Range Available

Pipette Model Volume Range (µL) Color Code Ordering Information
With stainless steel ejector With plastic Ejector
 P2L  0.2 - 2  Orange  FA10001M  FA10001P
 *P10L  1 - 10  Red  FA10002M  FA10002P
 P20L  2 - 20  Pale Yellow  FA10003M  FA10003P
 P100L  10 - 100  Salmon  FA10004M  FA10004P
 P200L  20 - 200  Yellow  FA10005M  FA10005P
 P1000L  100 - 1000  Blue  FA10006M  FA10006P
 P5000L  500 - 5000  Purple   **FA10007
 P10mL L  1 - 10 mL  Pale Blue   **FA10008

*P10L model can be used up to 0.5 µL
** No tip ejector required or needed